A Laser Power Meter consists of a display and a detector. The Laser Power Meters has been calibrated at a given wavelength before shipping.
1. Outline
Laser Power Meter using the amount of heat absorption measurement principle, thermal sensor measure the scope is very wide in spectral width and power, very high spatial uniformity, Without spot size, spot location or spot uneven impact. Can measure continuous wave laser output power, pulse repetition rate can also measure the average laser output power.
2. Technical parameters
1 Name: Laser Power Meter
2 Model: MT-92C
3 Range: 0-200W
4 Display: LED display
5 Minimum Resolution: 0.1W
6 Photosensitive area: 28mm
7 Spectral range: 8-11um
8 Calibration Wavelength: 10.6um
9 Response time: ≤5S
10 Cooling: Water-cooled
11 Applications: CO2 laser
12 Nonlinearity: ±5%
13. Accept the surface response nonuniformity:±5%
14.Consumption Power:≤8W
15.Power Requirements: AC 220V ± 10%, 50Hz
3. Usage
Please check whether the parts is complete after open the package. Accessories include: probe, base, display box, power cord, and product specifications.
2.Connecting signal lines and power cord ,open the display box on the power switch.
3. Power meter probe on cooling water, please.
4.According to the power of the test object, choose appropriate range through the switch.
5. The laser output should be vertical alignment probe center of power meter,and do power measurement.
6. Turn off the power supply when test is finished.
4. Notices
1.Handle with care, prevent precision components for loss or damage.
2.Pedestal of Power meter should be placed.
3.Power meter should be connected cooling water when working, long time testing ,pay attention to the temperature of water
4.Inspection or disassembling must cut off power supply firstly.
5.Long time no use, please pull out the power cord and signal lines.
6.If there are any problems please contact the manufacturer, it is strictly prohibited to disassemble.